==CALIBRATION== | ==CALIBRATION== | ||
The Pen should be calibrated if the batteries have been changed or if the LCD reads “CAL” followed by | *The Pen should be calibrated if the batteries have been changed or if the LCD reads “CAL” followed by | ||
“- - - -”. | *“- - - -”. | ||
Remember, calibration involves depressing the Operator Button TWICE and getting ONE | *Remember, calibration involves depressing the Operator Button TWICE and getting ONE | ||
line (- - - -). To take a reading requires depressing the Operator Button ONCE and getting | *line (- - - -). To take a reading requires depressing the Operator Button ONCE and getting | ||
TWO lines (====). | *TWO lines (====). | ||
1. Point the transducer end straight down towards the floor. | *1. Point the transducer end straight down towards the floor. | ||
2. Press the Operator Button “distinctly, two times, within 1.5 seconds.” | *2. Press the Operator Button “distinctly, two times, within 1.5 seconds.” | ||
3. Initially “CAL” appears on the LCD, followed by “UP” (after about 15 - 30 seconds!) | *3. Initially “CAL” appears on the LCD, followed by “UP” (after about 15 - 30 seconds!) | ||
4. Invert the transducer end to point straight up toward the sky. | *4. Invert the transducer end to point straight up toward the sky. | ||
5. A properly functioning Pen will display “”Good”. | *5. A properly functioning Pen will display “”Good”. | ||
If “bAd”, repeat the calibration again. | *If “bAd”, repeat the calibration again. | ||
If this is all unsuccessful, options include | *If this is all unsuccessful, options include | ||
cleaning the tip (compressed air is recommended) | *cleaning the tip (compressed air is recommended) | ||
change the batteries. Use the little probe that came with the Pen to open the | *change the batteries. Use the little probe that came with the Pen to open the | ||
battery compartment. | *battery compartment. | ||
Tech Support | *Tech Support | ||
www.reichert.com | *www.reichert.com | ||
www.tonopen.us | *www.tonopen.us | ||
1-888-849-8955 | *1-888-849-8955 | ||
[[Category:Procedures]] | [[Category:Procedures]] | ||
Revisión del 08:47 19 may 2016
CALIBRATION
- The Pen should be calibrated if the batteries have been changed or if the LCD reads “CAL” followed by
- “- - - -”.
- Remember, calibration involves depressing the Operator Button TWICE and getting ONE
- line (- - - -). To take a reading requires depressing the Operator Button ONCE and getting
- TWO lines (====).
- 1. Point the transducer end straight down towards the floor.
- 2. Press the Operator Button “distinctly, two times, within 1.5 seconds.”
- 3. Initially “CAL” appears on the LCD, followed by “UP” (after about 15 - 30 seconds!)
- 4. Invert the transducer end to point straight up toward the sky.
- 5. A properly functioning Pen will display “”Good”.
- If “bAd”, repeat the calibration again.
- If this is all unsuccessful, options include
- cleaning the tip (compressed air is recommended)
- change the batteries. Use the little probe that came with the Pen to open the
- battery compartment.
- Tech Support
- www.reichert.com
- www.tonopen.us
- 1-888-849-8955