Diferencia entre revisiones de «Tono-Pen use»

Sin resumen de edición
 
(No se muestran 6 ediciones intermedias de 2 usuarios)
Línea 1: Línea 1:
==To Use (if already calibrated==
==To Use (if already calibrated)==
*Depressing the Operator Button ONCE
*Depressing the Operator Button ONCE
*Monitor should show TWO lines (====).
*Monitor should show TWO lines (====).
Línea 15: Línea 15:
*#*If “bAd”, repeat the calibration again.
*#*If “bAd”, repeat the calibration again.


===If unsuccessful===
Options include:
*clean the tip (compressed air is recommended)
*change the batteries.
**Use the little probe that came with the Pen to open the battery compartment.
*Tech Support
**www.reichert.com
**www.tonopen.us
**1-888-849-8955


*If this is all unsuccessful, options include
==See Also==
**cleaning the tip (compressed air is recommended)
*[[Eye pain]]
**change the batteries. Use the little probe that came with the Pen to open the
*[[Acute angle closure glaucoma]]
**battery compartment.
*[[IOP]]
**Tech Support
[[Category:Procedures]]
***www.reichert.com
[[Category:Ophthalmology]]
***www.tonopen.us
***1-888-849-8955


[[Category:Procedures]]
==External Links==
 
===Videos===
{{#widget:YouTube|id=sBosdQ6o1oY}}

Revisión actual - 21:52 29 jun 2021

To Use (if already calibrated)

  • Depressing the Operator Button ONCE
  • Monitor should show TWO lines (====).
  • Use on eye

To Calibrate

  • The Pen should be calibrated if the batteries have been changed or if the LCD reads “CAL” followed by “- - - -”.
  • Calibration involves depressing the Operator Button TWICE and getting ONE line (- - - -).
  • If so,
    1. Point the transducer end straight down towards the floor.
    2. Press the Operator Button “distinctly, two times, within 1.5 seconds.”
    3. Initially “CAL” appears on the LCD, followed by “UP” (after about 15 - 30 seconds!)
    4. Invert the transducer end to point straight up toward the sky.
      • A properly functioning Pen will display “”Good”.
      • If “bAd”, repeat the calibration again.

If unsuccessful

Options include:

  • clean the tip (compressed air is recommended)
  • change the batteries.
    • Use the little probe that came with the Pen to open the battery compartment.
  • Tech Support
    • www.reichert.com
    • www.tonopen.us
    • 1-888-849-8955